Bridges
Bridge circuits measure small changes in impedance with great precision. Learn the Wheatstone bridge balance and output equations, quarter/half/full strain bridges, the Kelvin bridge for low resistance, and AC bridges for L and C.
A bridge is two voltage dividers side by side, with a detector between their midpoints. When the dividers have equal ratios, the midpoints are at the same voltage and the detector reads zero — the bridge is balanced. Any small change in one arm unbalances it, producing a small differential voltage centred on zero instead of a tiny change on top of a large one.
That makes bridges the standard way to measure strain, weight, pressure, temperature and component values with high sensitivity.
An analogy
A bridge is like a balanced see-saw: when both sides match it sits level; add a single grain to one side and it tips — you detect the tiny difference rather than weighing each side separately.
Balanced (null) measurement
In a classic laboratory Wheatstone bridge, you adjust a calibrated resistor until the galvanometer reads zero, then compute R_x from the known ratio. Because the detector only has to find zero, its own accuracy doesn't matter — only the reference resistors do.
Deflection bridges and strain gauges
In sensors the bridge isn't rebalanced; the output voltage is measured. Using more active gauges increases output and compensates temperature.
- Quarter bridge: one active gauge, V_o ≈ V_s·ΔR/(4R); slightly non-linear.
- Half bridge: two gauges (tension + compression or bending): 2× output, temperature-compensated.
- Full bridge: four active gauges: 4× output, linear and fully compensated — every load cell.
Kelvin double bridge
Below ~1 Ω, lead and contact resistance swamp the measurement. The Kelvin double bridge adds a second set of ratio arms so connection resistances cancel, measuring micro-ohms — busbar joints, shunts, motor windings.
AC bridges
With AC excitation and impedances, balance requires both magnitude and phase to match: Z₁Z_x = Z₂Z₃. Maxwell and Hay bridges measure inductance and Q; the Schering bridge measures capacitance and dielectric loss (tan δ) of high-voltage insulation; the Wien bridge measures frequency and is the basis of the Wien oscillator.
Ratiometric measurement
The bridge output is proportional to excitation, so modern designs use the same excitation as the ADC reference (ratiometric). Excitation drift then cancels exactly, and 24-bit sigma-delta ADCs with built-in PGAs read load cells directly.
Formulas
- Balance condition:
R₁ / R₂ = R₃ / R_x → R_x = R₂·R₃ / R₁ - Output (general):
V_o = V_s · [R_x/(R₃ + R_x) − R₂/(R₁ + R₂)] - Quarter bridge (small ΔR):
V_o ≈ V_s · (ΔR/R) / 4 = V_s·GF·ε/4 - Half / full bridge:
V_o ≈ V_s·GF·ε/2 / V_s·GF·ε
Worked example: Balanced Wheatstone bridge
Given: R₁ = 100 Ω, R₂ = 1000 Ω; R₃ = 470 Ω at balance
Find: R_x.
- R_x = R₂·R₃/R₁ = 1000 × 470/100 = 4700 Ω
Answer: 4.7 kΩ
Worked example: Quarter-bridge strain output
Given: V_s = 5 V; R = 350 Ω, ΔR = 0.35 Ω (500 µε, GF 2)
Find: Output voltage.
- V_o ≈ 5 × (0.35/350)/4 = 5 × 0.001/4 = 1.25 mV
Answer: ≈ 1.25 mV
Common mistakes
- Wrong: Measuring bridge output with a single-ended ADC referenced to ground. — Right: The output is differential around V_s/2; use a differential/instrumentation amplifier.
- Wrong: Using a quarter bridge without temperature compensation. — Right: Add a dummy gauge (half bridge) or use a full bridge so temperature effects cancel.
Applications
- Load cells & scales: Full-bridge load cells rated in mV/V are in every kitchen scale and industrial weighing system.
- Pressure sensors: Piezoresistive silicon diaphragms form full bridges on-chip.
- Insulation testing: Schering bridges and modern tan δ testers assess HV cable and transformer insulation ageing.
Key takeaways
- Balance: R₁/R₂ = R₃/R_x.
- Deflection output ≈ V_s·GF·ε·(n/4) for n active arms.
- Full bridges are linear and temperature-compensated.
- Kelvin bridges for micro-ohms; AC bridges for L, C and insulation loss.
Going further
- Six-wire load-cell connection: Separate sense wires measure excitation at the load cell itself, so voltage drop in long cables doesn't change the calibration.
- Bridge excitation choices: AC or chopped DC excitation removes thermoelectric offsets and 1/f noise; lower excitation reduces self-heating at the cost of signal.
- Automatic balancing bridges: Modern LCR meters are auto-balancing bridges: a feedback amplifier holds a virtual null while phase-sensitive detection separates real and imaginary impedance.