Measurement Techniques & Lab Instruments for Analog
Measuring analog circuits without fooling yourself. Learn DMM voltage, current and resistance measurement (input impedance, burden voltage, 4-wire ohms), oscilloscope bandwidth, sample rate, rise time and probe selection (10×, active, differential, current probes), probe loading and grounding (ground-spring vs long lead), triggering and measuring ripple and noise, function generators and their 50 Ω output, bench power supplies (current limit, remote sense), spectrum analysers and FFT, LCR meters, network/Bode analysers for loop gain, and measurement uncertainty — with Python quantifying probe loading, system rise time, ground-lead ringing and meter burden voltage.
Every instrument connected to a circuit becomes part of it: it loads nodes, adds capacitance and inductance, and has limited bandwidth and accuracy. Good measurements start with knowing how the instrument changes what you measure.
The goal is not just a number, but a number you trust — with an idea of its uncertainty.
An analogy
Measuring a circuit is like checking a tyre's pressure: every time you press the gauge on, a little air escapes. A good gauge (instrument) lets out so little that the reading still represents the tyre.
Multimeters
Use them well:
- Voltage: 10 MΩ input — fine for most nodes, not for 1 MΩ+ sources
- Current: series connection, burden voltage, fuse; prefer a shunt + voltage measurement
- Resistance: power off; 4-wire (Kelvin) for < 10 Ω
- True-RMS for non-sinusoidal AC; check bandwidth (often < 100 kHz)
- Diode test and continuity for quick fault finding
Oscilloscopes and probes
Choose bandwidth ≥ 3–5× the highest frequency of interest (or 0.35/t_r of the fastest edge × 3), sample rate ≥ 2.5–5× bandwidth, and enough memory to keep that rate over long captures. 10× passive probes must be compensated with the scope's calibrator square wave. Use differential probes for floating measurements (never float the scope's ground!), current probes for inductor and switch currents, and 1× probes or dedicated low-noise techniques for small ripple.
Signal sources and supplies
Function generators have a 50 Ω output: into a high-impedance load the actual voltage is double the display (unless set to 'High-Z'). Bench supplies should use current limiting when powering new boards and remote sensing for accurate voltage at the load. Electronic loads test regulators with steps.
Frequency-domain instruments
Spectrum analysers and scope FFTs show harmonics, spurs and noise floors (set RBW, window and averaging appropriately). Network/frequency-response analysers inject a signal and measure gain/phase — Bode plots of filters and loop gain of power supplies (with an injection transformer). LCR meters measure capacitors at a given frequency and DC bias, revealing ceramic derating and ESR. Always state measurement conditions and uncertainty.
Formulas
- Rise time ↔ bandwidth:
t_r ≈ 0.35/BW (Gaussian-like response) - Cascaded rise times:
t_r,displayed = √(t_r,scope² + t_r,probe² + t_r,signal²) - Probe loading:
|Z_probe| = R ∥ 1/(2πfC) - Ground-lead ringing:
f_ring = 1/(2π√(L_lead·C_probe)); L ≈ 10–15 nH/cm - Burden voltage:
V_burden = I·R_shunt - Loading error (DMM):
ε = R_source/(R_source + R_in)
Worked example: Probe impedance
Given: 12 pF at 100 MHz
Find: |X_C|.
- 1/(2π × 100 MHz × 12 pF)
Answer: ≈ 133 Ω
Worked example: Scope rise time
Given: 100 MHz bandwidth
Find: t_r.
- 0.35/100 MHz
Answer: 3.5 ns
Worked example: Cascaded rise time
Given: Scope 1.2 ns, signal 2 ns
Find: Displayed.
- √(1.2² + 2²)
Answer: ≈ 2.33 ns
Correct probing of a fast edge or ripple
WRONG: 15 cm ground clip → large loop → ringing + picked-up switching noise
probe tip ──── node
ground clip ─(15 cm loop)─ GND somewhere
RIGHT: ground spring (≤ 1 cm) at the capacitor terminals, 20 MHz bandwidth limit for ripple,
AC coupling, 1× or 10× as needed; for floating nodes use a differential probeThe ground lead's inductance and the probe's capacitance form a resonant circuit that rings at ~100 MHz and picks up magnetic fields from nearby switching loops. A short ground spring at the measurement point removes most of that artefact.
Probe loading, rise time, ringing and burden
import math
# Measurement: probe loading, system rise time, ground-lead ringing and meter burden
Cp = 12e-12
for f in (1e3, 1e6, 10e6, 100e6):
Xc = 1 / (2 * math.pi * f * Cp)
print(f"10x probe (10 Mohm || 12 pF) at {f/1e6:8.3f} MHz: |Z| ~ {1/math.sqrt((1/10e6)**2 + (1/Xc)**2)/1e3:10.2f} kohm")
tr_scope, tr_probe, tr_sig = 3.5e-9, 1.4e-9, 2e-9
tr_meas = math.sqrt(tr_scope**2 + tr_probe**2 + tr_sig**2)
print(f"100 MHz scope (3.5 ns) + probe: a 2 ns edge is displayed as {tr_meas*1e9:.2f} ns")
L_lead = 15e-9 * 15 # ~15 nH/cm, 15 cm ground clip
f_ring = 1 / (2 * math.pi * math.sqrt(L_lead * Cp))
print(f"15 cm ground lead ({L_lead*1e9:.0f} nH) with 12 pF rings at {f_ring/1e6:.0f} MHz -> use the ground spring")
# DMM on 200 mA range with 1 ohm shunt measuring a 150 mA load on a 3.3 V rail
print(f"burden voltage {0.15*1.0:.2f} V -> the DUT sees {3.3-0.15:.2f} V instead of 3.3 V")
# 10x probe (10 Mohm || 12 pF) at 0.001 MHz: |Z| ~ 7984.71 kohm
# 10x probe (10 Mohm || 12 pF) at 1.000 MHz: |Z| ~ 13.26 kohm
# 10x probe (10 Mohm || 12 pF) at 10.000 MHz: |Z| ~ 1.33 kohm
# 10x probe (10 Mohm || 12 pF) at 100.000 MHz: |Z| ~ 0.13 kohm
# 100 MHz scope (3.5 ns) + probe: a 2 ns edge is displayed as 4.27 ns
# 15 cm ground lead (225 nH) with 12 pF rings at 97 MHz -> use the ground spring
# burden voltage 0.15 V -> the DUT sees 3.15 V instead of 3.3 VA '10 MΩ' probe is only ~1.3 kΩ at 10 MHz because of its 12 pF — it can load high-impedance nodes badly. A 100 MHz scope shows a 2 ns edge as 4.3 ns, so fast edges need more bandwidth (≥ 3–5× the signal's). A 15 cm ground lead rings near 100 MHz. And a DMM's current shunt drops 0.15 V at 150 mA — enough to brown out a 3.3 V circuit under test.
Common mistakes
- Wrong: Floating the oscilloscope ground to measure mains-referenced nodes. — Right: Use a differential probe or isolated scope — floating is dangerous.
- Wrong: Uncompensated 10× probes. — Right: Compensate on the calibrator square wave.
- Wrong: Reading generator amplitude into high-Z as displayed. — Right: Set the generator load setting correctly.
Applications
- Power-supply ripple: Ground-spring, BW-limited measurement.
- Audio testing: THD+N analysers and FFTs.
- Loop-gain measurement: Injection with a network analyser.
- Component characterisation: LCR meter at bias.
Key takeaways
- Instruments load circuits — know R_in, C_in, burden.
- Bandwidth ≥ 3–5× signal; t_r = 0.35/BW.
- Short ground connections for fast signals and ripple.
- Use the right tool: DMM, scope, FFT, LCR, network analyser.
Going further
- Measurement uncertainty budgets: Traceability and GUM.
- Power-rail probes: Offset and low-noise probes.
- Time-domain reflectometry: Cables and PCB impedance.